Variation of the count-dependent Metz filter with imaging system modulation transfer function.

Research paper by M A MA King, R B RB Schwinger, B C BC Penney

Indexed on: 01 Mar '86Published on: 01 Mar '86Published in: Medical physics



Abstract

A systematic investigation was conducted of how a number of parameters which alter the system modulation transfer function (MTF) influence the count-dependent Metz filter. Since restoration filters are most effective at those frequencies where the object power spectrum dominates that of the noise, it was observed that parameters which significantly degrade the MTF at low spatial frequencies strongly influence the formation of the Metz filter. Thus the radionuclide imaged and the depth of the source in a scattering medium had the most influence. This is because they alter the relative amount of scattered radiation being imaged. For low-energy photon emitters, the collimator employed and the distance from the collimator were found to have less of an influence but still to be significant. These cause alterations in the MTF which are more gradual, and hence are most pronounced at mid to high spatial frequencies. As long as adequate spatial sampling is employed, the Metz filter was determined to be independent of the exact size of the sampling bin width, to a first approximation. For planar and single photon emission computed tomographic (SPECT) imaging, it is shown that two-dimensional filtering with the Metz filter optimized for the imaging conditions is able to deconvolve scatter and other causes of spatial resolution loss while diminishing noise, all in a balanced manner.